Operating Parameters

Method: Ion selective electrode (ISE) Method

Code: NW-S800

Sample: Whole blood, serum, plasma and diluting urine

Sample Volume: 150ul

Measuring Range and Accuracy

 Electrode

Linearity

Resolution

CV

Potassium ion(K)

0.50-10.00mmol/L

0.01mmol/L

<1%

Sodium ion (Na+.)

80.0-200.0mmol/L

0.1 mmol/L

<1%

Chloride ion (Cl-.)

50.0-200.0mmol/L

0.1 mmol/L

<1%

Ionized calcium (Ca++.)

0.10-50mmol/L

0.01 mmol/L

<2%

pH

4.00-9.00

0.01

<1%

Lithium ion (Li+)

0.3-3.00mmo/L

0.01

<2%

Rinsing Method: Quickly rinsing solenoid valve, cleaning thoroughly electrode pipe, no blocking.

Measuring Time: Less than 50S (depends on electrode in stability)

Display: 5.6 inch Color Display  touch screen.

Language: English and other language as required.

Sensor: Intelligent Sampling system, full automatic taking sample.

Reagent: can do with reagent bottle or Reagent Pack (both)

Printer: Built-in thermal printer

Interface: Standard RS232 interface, connect to work station and LIS system

Temperature: 5-40 ℃, less than 85% relative humidity

Power Requirements: 220V±22V,50Hz±1Hz